Cooling capacity (W) 1300/1400
Heating capacity (W) 230 / 350
Circulating water temperature range (℃) 5 to 40
Temperature stability (℃) ±0.1
Circulating water rated flow rate (ℓ/min) 15/15
Cooling scheme Air-cooling type
Cooling capacity (W) *1 1300 / 1400
Heating capacity (W) *1 230 / 350
Circulated water *2 Clean water, Ethylene glycol solution (concentration 15 wt% or less), Propylene glycol solution (concentration 15 wt% or less)
Circulating water temperature range (℃) 5 to 40
Temperature stability (℃)*3 ±0.1
Use Area Temperature (℃) 5~45
Circulating water rated flow rate (ℓ/min) 15/15
Maximum discharge pump head (m) *4 14/20
Tank capacity (ℓ) Approx. 10
Dimensions (mm) *5 H610×W400×D445
Coating Color Munsell N-87
Weight (kg) Approx. 42
Casters (Φmm×unit) 38×4
Rated Voltage (V) *6 1-phase 200 VAC
Power Consumption (kW) *1 0.60 / 0.73
Current consumption (A) *1 3.5 / 3.7
Recommended breaker capacity (A) 10
Circulating water inlet / outlet port size Rc1/2
Drain discharge port size G1/2
Components / Compressor Fully sealed rotary type
Components / Condenser Fin and tube type
Components / Evaporator Structure : Plate-type heat exchanger
Material : SUS316 (copper brazing)
Components / Circulating pump Structure : Magnetic centrifugal pump
Output : 0.180/0.216 (kW)
Certification CE / ETL
Refrigerator R1234yf 300g
Noise (dB) *7 51 / 54
Safety circuit / Compressor Thermal protector
Safety circuit/Pump Thermal protector
Safety circuit / Fan motor Thermal protector, Current fuse
Input / Alarm output No voltage open-contact (Max. AC250V/1A, DC30V/2A)
Input/Output / Water temperature upper / lower limit output No voltage open-contact (Max. AC250V/1A, DC30V/2A)
Input/Output / Operation stop output No voltage open-contact (Max. AC250V/1A, DC30V/2A)
Input/Output / Pump output No voltage open-contact (Max. AC250V/1A, DC30V/2A)
Input/Output / Emergency stop input No-voltage input
Input/Output / Operation stop input No-voltage input
Input/Output / Error reset input No-voltage input
Input/Output / External temperature sensor *8 Pt100
Input/Output / Ethernet 10BASE-T/100BASE-TX
Socket(TCP/IP)、MC protocol(UDP/IP)、Modbus TCP
Display / operation method Monochrome display, key input
Log data Recorded items:
Date and time, error number, circulation water set temperature, circulation water outlet temperature, external temperature, chiller run/stop, pump/fan/compressor run/stop, pump independent run status, emergency stop command input terminal status, run-stop command input terminal status, error reset command input terminal status, false warning output status, compressor high-pressure pressure, compressor low-pressure pressure, compressor ejection temperature, compressor intake temperature, water empty sensor status, fan control status, coolant operation volume, high-pressure protection control status, low-pressure protection control status, overheating protection control status, freezing protection control status.
Recording period: Approx. 1 year and 6 months
Recording interval *9: 2 minute intervals
Export method: USB memory, Socket communication
Functions Graph display, chiller operation status display, error history display, operation time display, calendar, write memo, water temperature error settings, operation timer settings, password settings, key lock, freeze protection operation, log data USB memory write, external temperature sensor input, cascade control