0865 735 110

Measuring methodSpeedmeter/Counter [Switchable] SPD-HD

Name
Model
Measuring method
Speedmeter/Counter [Switchable] SPD-HD
PeriomaticTMB method Cycle calculation method

Input Number of input points
Input frequency
Input resolution
Input Signal
Input method
1 point
0.0006Hz to 1.0MHz
Approx. 10.4ns (96MHz)
5V logic, zero cross, NPN open collector, 12V logic, input customized signal
Single-phase signal

[*] General-purpose
input option
Input Level and Sensitivity
5V logic

Zero-cross
NPN open collector
12V logic
Customized input
Trigger level
2.5V
0V
2.5V
6V
0-9.0V
in 10mV units
Hysteresis
±0.4V
±0.03V
±0.4V
±0.4V
[Small]±0.03V, [Medium] 0.4V,
[Large]±1.3V, trigger level 4.7V or lower/
±2.6V, trigger level 4.7V or higher
 Input pulse
Input resistance

Input withstand voltage
[Small] [Medium] 0.2 μs or more   [Large] 0.3 μs or more, for both H and L levels
5V logic, zero cross, NPN open collector, and Input customized signal (trigger level 4.7V or lower): 10kΩ
12V logic, input customized signal (trigger level 4.71V or higher): 5kΩ
±30V  *Inputs and outputs are not insulated for the general-purpose input option.

[D] Line driver
input option
Input signal
Input withstand voltage
Recommended line driver
Termination resistance
Line driver signal
±25V (differential voltage)
Equivalent to AM26LS31
300Ω (standard specification)  *Input and output are isolated for the line driver input option.

[W] 2-Phase(A/B)
input option
Input Level and Sensitivity
5V logic

Zero-cross
NPN open collector
12V logic
Customized input
Trigger level
2.5V
0V
2.5V
6V
0 / 2.5 / 6V
Hysteresis
±0.4V
±0.03V
±0.4V
±0.4V
Small:±0.03V, Medium : 0.4V

Input low-pass filter
Trigger edge
Input withstand voltage
Input connector

Sensor Power supply output[Option]Select from OFF, 500Hz, 5kHz, 120kHz, or 800kHz
Fall
±30V
[Retractable BNC] Single-phase input, analog output
[Screw Terminal Plug Connector] 2 sensor power supply points, input signal, analog output
[M8 connector] 1 sensor power supply, input signal, analog output
[H] DC +12V±5% 60mA max / [L] DC +5V±5% 150mA max
*For M8 connector type, either 5V or 12V is specified.

CalculationMeasurement mode
Calculation rate
Frequency divider ratio
Output moving average
Dynamic PredictiveTM
Control input
[Setting contents] [Operation Method] Low cut function
Velocity meter/Frequency meter/Period meter (seconds only) /Counter
Set the reference pulse number, reference change, and speed unit time to the input frequency.
1 to 60000
1 to 99
8 levels (including continuous and stop prediction)
None or 1 point (depending on model)
Functions: hold (current/max./min. value, max. fluctuation range), Reset (Count) can be set
ON when control input (CTL*) pin is shorted to GND; setting is reflected when open.
Low cut function Stop judgment at a speed lower than the set value

DisplayIndicator
Display range
Decimal point position
Display accuracy
Display update time
Power ON/OFF indication
16 characters, 2 lines, character height 6mm OLDE
[Single phase]: 0.00000 – 999999     [Phase A/B]: ±9999999
Set in program mode, digits 0 – 5
Within 20ppm ±1 count
0.1 – 9.9s, settable in 0.1s increments
Lit when LED with built-in power switch is ON

USB
communication
Communication specification
Communication function
Output update time
USB2.0 (Serial port communication by USB virtual COM 230.4kbps)
Writing/reading of set values/continuous output of measured values
1 to 9999.9ms, settable in 0.1ms increments (No output when output update time is set to 1 to 9)

General
Information
Operating temp./humidity
Ambient operating temp.
Power supply voltage

Power consumption
Isolation
Withstand voltage
Dimensions
Accessories-10°C to +50°C / 35 to 85%RH (non-condensing)
-20 °C to +65 °C
[Standard]: AA battery 1.5V x 4 (DC6V), usable for approx. 8 hours (With no sensor power output)
[AC adapter] Use the supplied AC adapter.
3W or less (Sensor power output is not included)
Power supply / Other input/output
Power supply / Other input/output: 500V DC between each terminal for 1 minute
94 W×32 H×150 D [mm] AC adapter, carrying case, 4 AA batteries (cable not included)

Analog
output option
Output signals
D/A conversion method
Output resolution
Load resistance
Output accuracy
Temperature fluctuation
Output scaling
Output update time
Input/output delay time
[E1] 0 to 10 V, [E5] 0 to 5 V, [R1] ±10 V, [R5] ±5 V, [I1] 4 to 20 mA    [Select a model]
DAC conversion method
16bit (58000 or more)
[Voltage output] 4.7kΩ or more / [Current output] 300Ω or less
[Voltage output] ±0.1% of FS @ 23°C / [Current output] ±0.1% of FS @ 23°C
±200ppm/℃ or less
Arbitrary scaling is possible by full scale and zero scale settings
0.1 to 9999.9ms settable in 0.1ms increments
Output update time + analog rise time 150 μs max.

CAN OptionCommunication specifications
Communication function
Setting item
Output update timeCAN2.0B 10 k / 20 k / 50 k / 100 k / 250 k / 500 k / 1M bps [Format] Intel/Motorola [Frame] 11bit/29bit
Writing/reading of set values/continuous output of measured values
CAN transmit ID, CAN receive ID
1 to 9999.9 ms (settable in 0.1 ms increments)
(When output update time is set to 1 to 9, instantaneous values are output every 1 ms.)

Ethernet
communication
option
Connector
Interface
Communication specification
Output update time
Transmission and Communication Compatibility
LED indication
Link/Activity displayRJ-45 (network interface)
100BASE-TX/10BASE-T (automatic recognition) Transmission speed standard: Category 5 / Category 4

Protocol : TCP/IP method Connected by Telnet communication (serial to Ethernet conversion)
1ms – 9999.9ms (Output update time is 1ms or more. No output when set to 1 to 9)
DIX Ethernet Version 2.0 / IEEE 802.3

100BASE-TX / 10BASE-T
Full duplex / half duplex (Telnet cannot handle communication networks encrypted by SSL/TLS or SSH)

.
.
.
.