0865 735 110

Variable-area flow meter Tokyo Keiso MA-920

 

Flow Measuring Instruments
Variable-area flow meter

Metal tube area flow meter for small flow rates

The MA-920 is a small area flowmeter that measures and outputs minute flow rates of various liquids and gases. It is an advanced microflow meter that combines many years of manufacturing know-how of area flow meters with state-of-the-art electronics technology.
In the past, there were various difficulties in improving the accuracy of area flow meters for micro-flow rates, such as machining accuracy, friction acting on floats, and non-linear characteristics.
In the MA-920, these error factors are automatically corrected by a microprocessor equipped with calibration data, enabling high-precision measurement in the minute flow range. The 2-wire DC4 – 20mA wiring scheme makes it extremely easy to construct on site.

Catalogs & Pamphlets

* The information is subject to change without notice due to product improvement.

forte

  • Covers a small flow range with a minimum flow rate of 0.6 ~ 3 L/h (water)
  • Magnetic sensor detection method eliminates hysteresis, achieves high accuracy and high reproducibility
  • Easy-to-read LCD digital display
  • 2-wire, DC 4 ~ 20 mA output
  • Exd II CT6, explosion-proof structure for hydrogen zero encirclement. There is no problem using it in hazardous areas
  • No damper required even in gas specifications

Standard Specifications

Measurement target Liquids in general, gases in general (not suitable for liquids containing solids)

However
Caliber (meter size) 1/2: Viscosity up to 2.0 mPa・s
Caliber (meter size) 3/4, 1: Viscosity up to 5.0 mPa・s

caliber 1/4, 3/8, 1/2, 3/4, 1
10A, 15A, 20A, 25A
(meter size) 1/2, 3/4, 1
Flow range
Liquid measurement (water equivalent) * The flow rate range varies depending on the viscosity.
Max. 60 ~ 600 L/h Min. 0.6 ~ 3 L/h

Gas measurement (air equivalent)
Max. 2.2 ~ 22 m³/h (nor) Min. 10 ~ 100 L/h (nor)

Fluid pressure 2.94 MPa Max. for
high pressure 19.6 MPa Max. (Body material is SUS316) In
the case of flange connection, within the flange standard
Process Connections RC screws, JIS10KFF flanges
(optional) Various screws such as NPT, non-JIS10KFF flanges
Measurement accuracy Full scale 10 L/h (water) or more ±2% F.S.
Full scale less than 10 L/h (water) ±3% F.S.
Full scale 100 L/h (nor) (air) or more ±2% F.S.

Contact usInquiries about products

We respond to various inquiries such as inquiries about products and technical questions.
.
.
.
.